PCR
Items 51 to 79 of 79
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 700.00 (ex. VAT) 1 in stock
Corbett Research Rotor-Gene RG-3000 Lab Spares/Repairs
ITEM-25811
This was removed from a university lab where it was surplus to requirements.
The unit powers up but when we tried to connect it to a PC with software we had some issues. (Software not included).
We installed software on a test PC and it did seem to connect to the RG3000, but it is stating that there is a Fatal Hardware Malfunction. The machine can't be used until the distributor has service it. After starting it again we got another error: Thermistor is Open.
We have removed the casing to check inside the unit and cannot find any obvious issues.
There is an LCD panel inside that is showing 'TEMP OVER 140/127.5/2.1/152.0'. This displays as soon as the unit is switched on, so it looks like there is some sort of temperature error happening. There are 4 x PTC EPCOS C810 Thermistors in the power supply section of the unit. It's possible that there is a fault with at least one of those.
It is in otherwise reasonable cosmetic condition with marks from use, but is sold for spares/repairs
The item holds up to 72 x 0.1ml strip tubes.
£ 600.00 (ex. VAT) 1 in stock
Applied Biosciences 7900 HT Fast Real-Time PCR Complete System Lab
ITEM-24006
Applied Biosciences 7900 HT Fast Real-Time PCR Complete System
This was removed from a university lab where it was surplus to requirements.
We were able to get the unit to communicate with the PC but we were unable to get it to run cycles due to a serial number conflict. We were able to open the door but not close it again.
Included with it are;
1 x DELL PC
Windows Vista Business
1 x SDS 2.3
Item is in a good used condition. The Item powered on and working, but not connecting to PCR Tester
1 x Bti PRGF Desktop Centrifuge
It is in good used condition. The item powered on and was tested, all working
4 x PGRF CF10142 LOT 2894
1 x Diagnostica Stago Printer
Item powered on and tested, all working
1 x Datalogic QS6000 Plus Barcode Reader
The item powered on and was tested, all working
Associated books and a spare Block Assembly.
£ 600.00 (ex. VAT) 1 in stock
Bio-Rad Tetrad 2 Peltier Thermal Cycler Lab
ITEM-44522
Bio-Rad Tetrad 2 Peltier Theramal Cycle Lab
This item was removed from a university where it was surplus to requirement.
The unit powers up and is in good working order (thermometer not included).
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Roche LightCycler 480 II/96 PCR Real Time Thermal Cycler
ITEM-45158
Roche LightCycler 480 II/96 PCR Real-Time Thermal Cycler
This item was removed from a university where it was surplus to requirements.
The unit powers on and is in good used cosmetic condition.
During setup with software, the unit failed to initialise, displaying the following errors:
- Error 100900270: Restart instrument, inform Roche service.
- Warning 100600130: Command not allowed.
Due to the errors faced when tested, this item is listed as Spares/Repair.
Specifications:
- Excitation Filters Installed:
- 440, 465, 498, 533, 618
- Emission Filters Installed:
- 488, 510, 580, 610, 640, 660
- Xenon Lamp:
- Absolute Intensity: 9012
- Operation Time: 865 hours
- Block Type: 384 Silver Block
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Applied Biosystems Veriti 96 Well Thermal Cycler Lab Spares/Repairs
ITEM-41561
Applied Biosystems Veriti 96 Well Thermal Cycler Lab Spares/Repairs
This item was removed from a university where it was surplus to requirement.
It is in good cosmetic condition but does not power on, therefore, it is listed as Spares/Repair.
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Geneflow Sensoquest LabCycler Gradient Laboratory Thermal Cycler Lab Faulty
ITEM-41560
Geneflow Sensoquest LabCycler Gradient Laboratory Thermal Cycler Lab
This item was removed from a university where it was surplus to requirement.
The unit did not power up in testing. it is sold for spares/repairs
£ 500.00 (ex. VAT) 1 in stock
G-STORM Thermal Cycler GS4 GS00004 Lab
ITEM-41152
This was removed from a university lab where it was surplus to requirements.
The unit powers up and is in full working order.
It is in good cosmetic condition with minor marks.
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Eppendorf Realplex 4 MasterCycler Epgradient S Real Time Cycler 96-Well Lab
ITEM-31867
Eppendorf Realplex 4 MasterCycler Epgradient S Real Time Cycler 96-Well
This item was removed from a university where it was surplus to requirement.
It is in good cosmetic condition and powers on, however it does not include the required controller. We made several attempts to connect it to the software from another unit (not included), but it would not communicate Therefore, it is listed as Spares/Repair.
Includes an Eppendorf USB drive with pre-installed software and a digital user manual.
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Roche LightCycler 480 II /96 PCR Real Time Thermal Cycler Lab RT-PCR Faulty
ITEM-20612
This unit was removed from a lab as surplus to requirements. The lab assured us it was in working order when last used, but we do not have a PC or software to test this instrument at our site.
In the testing we were able to carry out, the item powers on, both front lights were flashing orange.
We attempted to test it with a PC and software (not included). The unit is failing to initialise during the startup. It seems to be failing at the 'camera cooling surveillance' section.
Item is in a good used cosmetic condition.
It is sold for spares/repairs
SPARES/REPAIR ITEM
SPARES/REPAIR ITEM
£ 500.00 (ex. VAT) 1 in stock
Roche LightCycler 480 II/96 PCR Real Time RT-PCR Machine Lab Spares/Repairs
ITEM-36517
Roche LightCycler 480 II/96 PCR Real Time RT-PCR Machine
This was removed from a university lab where it was surplus to requirements.
The unit powers up but the LightCycler is showing errors during the installation
100900270 - Restart instrument.
100600130 - Command not allowed - I believe this error is actually caused by the one above.
We have added images of the errors etc, but the computer will not be included.
Lamp Absolute Intensity - 9853
Operation Time - 572h.
It is in good cosmetic condition with minor marks from use.
It is sold for spares/repairs.
It is supplied with the following accessories.
1x ThermoCycler 96-s (boxed)
S/N: 030203
P/N: 05015162001
1x ThermoCycler 384 silber (fitted)
S/N: (21) 7298
P/N: 050154001
1x LightCycler 480 Software CD Release 1.5.0
7x Boxes of 96-well PCR microplates
O/N: I1402-9909
13x Boxes of 384-well PCR plates
O/N: E1402-9909
3x 2.5a 20 mm anti surge fuses
1x Maxon Motor (suspected faulty, as mentioned in report)
5x Mains power leads
1x 24 volt power supply
£ 450.00 (ex. VAT) 1 in stock
Techne TC-4000 Thermal Cycler Lab
ITEM-36564
Techne TC-4000 Thermal Cycle
This was removed from a university lab where it as surplus to requirements.
The unit powers up and is in good working order.
It is in good cosmetic condition with minor marks.
Thermometer used for testing is not included.
£ 385.00 (ex. VAT) 1 in stock
Techne Progene Thermal Cycler Model FPR0GO5D Lab
ITEM-35942
Techne Progene Thermal Cycler Model FPR0GO5D
This was removed from a university lab where it was surplus to requirements.
The unit powers up and is in good working order. (Digital thermometer not included).
It is in good cosmetic condition with minor marks.
£ 360.00 (ex. VAT) 1 in stock
Techne Thermal Cycler 3Prime with 24 x 0.1ml Well Block Lab
ITEM-26484
This Techne Thermal Cycler 3 Prime was removed from a university lab where it was surplus to requirements.
The item is in reasonable cosmetic condition. There are stains on the top of the lid and by the LCD display. The right hinge has broken on the inside, which causes that section to not sit flat.
The rubber seal around the heat block has broken and can come loose. This doesn't seem to affect its ability to retain heat. The 24x well block fits 0.1ml test vials.
The item powered on without issue. It was tested with a digital thermometer (not included).
When set to 37 the thermometer measured 36.7˚C.
When set to the max of 100 the thermometer measured 100.0˚C.
The heated lid works as intended.
£ 360.00 (ex. VAT) 1 in stock
Corbett CG-1 Palm Cycler Laboratory Thermal Cycler PCR Machine Lab
ITEM-20214
This Corbett CG-1 Palm Cycler PCR Machine was removed from a university lab where it was surplus to requirements.
The unit powers up and is in very good cosmetic condition.
We had the unit inspected by a service engineer. They have checked the system and recovered the Palm Cycler software. They ran the system through steps of range temperatures and time slots. The system runs as required within the time frame of each temperature. They inserted an external temperature probe while running the experiment. The measured temperature was close to what was set.
It is in full working order.
Block Spec:
96 well block 96x 0.2ml tubes (read from manual)
Dia 4.4mm x Depth9.6mm
£ 340.00 (ex. VAT) 2 in stock
Techne Flexigene FFG02FSD Benchtop Thermal Cycler PCR Lab
ITEM-14889
This Techne Flexigene FFG02FSD Bench-Top Thermal Cycler was removed from a laboratory where it was surplus to requirements.
The unit is in good condition but is showing signs of wear. It powers on, and a basic program was run to test the lid and the sample tray heater, which is working. It has not been tested further.
£ 320.00 (ex. VAT) 1 in stock
Applied Biosystems ABI Prism 7700 Sequence Detection System Model - 7700 Lab
ITEM-26200
This Applied Biosystems ABI Prism 7700 Sequence Detection System Model - 7700 was removed from a university lab where it was surplus to requirements.
The unit powers up but we were unable to test it further.
It is in reasonable cosmetic condition but has marks from use.
Pre-installation Manual
Planned Maintenance Protocol
Site Preparation & Safety Guide
£ 310.00 (ex. VAT) No longer available
Bio-Rad DNAEngine Peltier Thermal Cycler Model PCT0200 Lab
ITEM-38687
This Bio-Rad DNAEngine Peltier Thermal Cycler Model PCT0200 was removed from a university lab where it was surplus to requirements.
The unit powers up and is in good working order. It is in good cosmetic condition with minor marks.
It is fitted with a 96 well block.
£ 300.00 (ex. VAT) 1 in stock
Eppendorf Mastercycler Epgradient 6320 Thermal Cycler & Controller
ITEM-17358
This Eppendorf Mastercycler Epgradient Thermal Cycler with Controller was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested with a digital thermometer (not included). It is in full working order.
It is in good cosmetic condition with minor marks from use.
£ 265.00 (ex. VAT) 1 in stock
Bio-Rad Dyad Disciple Peltier Thermal Cycler PTC0221 2-48 Well & 96 Well Block
ITEM-36926
Bio-Rad Dyad Disciple Peltier Thermal Cycler PTC0221 2-48 Well & 96 Well Block
This was removed from a university lab where it was surplus to requirements.
The unit powers up but we were unable to test it without the control unit.
Bio-Rad Dyad Disciple Peltier Thermal Cycler
Model - PTC0221
SN - DY005562
2-48 Well Alpha Unit Block Assembly
SN - AL037212
96 Well Alpha Unit Block Assembly (ALP2296)
SN - AL110470
Holds 0.2ml vials
£ 265.00 (ex. VAT) 1 in stock
Biometra UNO II Thermoblock Thermal Cycler w/ 48 Block for 0.2ml Tubes Lab
ITEM-20719
Biometra UNO II Thermoblock Thermal Cycler w/ 48 Block for 0.2ml Tubes
This was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested with a digital thermometer (not included).
When set to 37˚C the display showed 37˚C and the thermometer measured 36.5˚C.
When set to 99.9˚C the display showed 100˚C and the thermometer measured 99.4˚C.
When set to 0˚C the display showed -1˚C and the thermometer measured -1˚C.
It seems to be in working order.
It is in reasonably good cosmetic condition with some marks from use.
£ 260.00 (ex. VAT) 1 in stock
G-Storm GS1 Thermal Cycler GS00001 96 x 0.1ml Lab
ITEM-32079
This G-Storm GS1 Thermal Cycler GS00001 96 x 0.1ml was removed from a university lab where it was surplus to requirements.
The unit powers up, and is in good working order.
It is in good cosmetic condition with minor marks from use.
£ 255.00 (ex. VAT) 1 in stock
MJ Research DNA Engine PTC-200 Thermal Cycler Lab
ITEM-25988
This MJ Research DNA Engine PTC-200 Thermal Cycler was removed from a university lab where it was surplus to requirements.
The unit is in full working order. It comes with a 384 well block which has been tested at 37ºC and 99ºC.
It is in reasonably good cosmetic condition with some minor marks from use.
£ 250.00 (ex. VAT) 1 in stock
MJ Research DNA Engine Opticon 2 CFD-3220 System w/ Software Lab
ITEM-44987
MJ Research DNA Engine Opticon 2 CFD-3220
This system was removed from a university lab where it was surplus to requirements.
The unit is in good cosmetic condition and powers on. It successfully connects to software and a PC, and programs run as expected.
Software:
- Opticon 2 - Version 3.0
Temperature Testing:
- Tested at 37.0ºC, 105.0ºC, and back to 0.0ºC
£ 250.00 (ex. VAT) 1 in stock
Labtech G-storm GS0001 Laboratory Thermal Cycler Lab
ITEM-34844
This Labtech G-storm GS0001 Laboratory Thermal Cycler was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested through cycles. It is in good working order.
It is in good cosmetic condition with marks from use.
£ 195.00 (ex. VAT) 1 in stock
Techne FlexGene FFG02FSD Benchtop Thermal Cycler Lab
ITEM-20281
This Techne FlexGene FFG02FSD Benchtop Thermal Cycler was removed from a university lab where it was surplus to requirements.
The unit powers up and has had basic temperature testing carried out with a digital thermometer (not included). We have not run any cycles.
When set to 37˚C the thermometer measured 37.1˚C
When set to 99˚C the thermometer measured 101.1˚C
The heated lid function works as intended.
The lid's max temperature is 110˚C
There are prior saved programs that cannot be deleted without the password.
The information menu states that the unit has run:
Unit cycles - 20178
Block cycles - 18896
It is in good cosmetic condition with some marks from use.
£ 185.00 (ex. VAT) 1 in stock
Helena Biosciences Cycler HBCY052 60 Well Thermal Cycler Lab
ITEM-18521
This Helena Biosciences Cycler HBCY052 60 Well Thermal Cycler was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested purely for heating purposes against a digital thermometer (not included). We have not run any cycles with it. When tested to 37 degrees it held successfully and when tested to 94 degrees it reached 81 degrees. The tube wells are 5mm in diameter & roughly 11.4mm deep/ We did not test it any further.
It is in good cosmetic condition with some light marks from use.
£ 180.00 (ex. VAT) No longer available
MJ Research PTC-100 Programmable Thermal Controller Cycler Lab
ITEM-26486
This MJ Research PTC-100 Programmable Thermal Controller Cycler was removed from a university lab where it was surplus to requirements.
The item is in reasonable cosmetic condition with scratches and marks.
The 60x well block fits 0.5ml test vials.
The item powered on without issue and the heated lid works. It was tested with a digital thermometer (not included).
When set to 37˚C the thermometer measured 37.0˚C.
When set to 100˚C the thermometer measured 101.9˚C.
It comes with an HBA-1600 hot bonnet heated lid.
£ 160.00 (ex. VAT) 1 in stock
Perkin Elmer DNA Thermal Cycler Lab
ITEM-31322
This Perkin Elmer DNA Thermal Cycler was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested at multiple temperatures (digital thermometer not included). It is in full working order.
It is in reasonably good cosmetic condition but has some marks from use.
Block Holds 48 x 0.5ml
£ POA 1 in stock
Quanta Biotech QTAS TAS16 Temperature Acquisition System Thermal Cycler Analysis
ITEM-45230
This item is new in stock. Please enquire for further details.
Items 51 to 79 of 79