LMS Cooled Refrigerated Laboratory Incubator Model - 305 NP Lab
Item Code: ITEM-26190
Condition: Used
Brand: LMS
Location: Wales
£ POA
SOLD - Unfortunately this item is no longer available. However, we may have similar items available for sale. Please get in touch with us at , or by phone on +44 29 2086 4661.
Description
This LMS Cooled Laboratory Incubator Model - 305 was removed from a university lab where it was surplus to requirements.
The unit powers up and was tested with a digital thermometer (not included).
It is in reasonably good cosmetic condition with some marks from use.
It was cooled down to 4˚C which it reached after about 15 minutes. it was then set to -10˚C. It reached -9˚C after a couple of hours.
It was then set to 37˚C which it reached in 10 minutes and then 50˚C which it reached in a further 10 minutes.
Inside dimensions - D20-28 W44 H53cm
Upper shelf - 27 x 43cm
Lower shelf - 13 x 4cm
Item Details
Property | Value | Tested value |
---|---|---|
Part / Catalogue Number | 305 | |
Type | Floorstanding | |
Heating Method | Unspecified | |
Refrigeration | Yes | |
Refrigerant | Yes | |
Minimum Temperature (ºC) | -10 | |
Maximum Temperature | 50 | |
CO2 Sensor Type | Unspecified | |
With Hot Air Sterilisation | No | |
Humidifier | N/A | |
With UV Lamp | No | |
Physical Specifications | ||
Interior Volume (L) | 60 |
Delivery
Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.
Additional terms and conditions
Incubators may be transported via a range of delivery methods, including palletised courier networks or direct delivery to the intended location of use. During transit, internal sensors, fans, or control systems may be affected by vibration or movement.
To ensure optimal performance and accurate environmental conditions, it is essential that all incubators are allowed to stabilise upon arrival and are professionally calibrated before first use. Calibration ensures the unit reaches and maintains the desired temperature, humidity, or CO₂ levels reliably.
Please ensure:
- The unit is left for a minimum of 12 hours upright and unpowered after delivery.
- Calibration is completed in accordance with manufacturer specifications before any use.
Failure to calibrate prior to use may result in inaccurate conditions and affect experimental integrity or compliance with laboratory standards.
Return policy
Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.
NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.