This RS Biotech Galaxy R CO2 Incubator Model - 170-300 was removed from a university lab where it was surplus to requirements.
The unit powers up and has been tested with 5% CO2 and a digital thermometer (not included).
When set to 37˚C the display showed 36.9˚C and the thermometer measured 37.4˚C. It took about 3 hours.
When set to 50˚C (max) the display showed 49.8˚C and the thermometer measured 50.6˚C. It took about 6 hours to get there.
The CO2 level climbed from it's setting of 5% up to 7%.
It is in reasonably good cosmetic condition but has marks from use.
Inside dimensions - D56 W52 H56cm
1 Shelf - 52 x 48cm