This Gallenkamp Plus II Laboratory Incubator Cat No. IPL.075.FD1.C was removed from a university lab where it was surplus to requirements.
The item is in reasonable cosmetic condition but has marks and stains from use.
The unit powers on fine and the interior shaker works as intended however the higher it is set on the dial the slower it goes and vice versa. The shaker cannot be operated at temperatures higher than 60˚C.
The unit was tested with a digital thermometer (not included)
When set to 37˚C the display showed 35.6˚C and the thermometer 37.3˚C.
When set to 60˚C the display showed 56.8˚C and the thermometer 58.7˚C.
We then removed the shaker and the oven was set to 108˚C, which is the highest we could get it to go on the dial.
It reached 98.6 according to the display, while the thermometer measured 104.0˚C.
The shaker is a Gallenkamp Shaker Platform Cat no. IPL000.0S1.9.
Item Details
Property | Value | Tested value |
---|---|---|
Type | Unspecified | |
Refrigeration | No | |
CO2 Sensor Type | Unspecified | |
Heating Method | Unspecified | |
With Hot Air Sterilisation | Unspecified | |
Humidifier | Unspecified | |
With UV Lamp | Unspecified |
Delivery
Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.
Additional terms and conditions
Incubators may be transported via a range of delivery methods, including palletised courier networks or direct delivery to the intended location of use. During transit, internal sensors, fans, or control systems may be affected by vibration or movement.
To ensure optimal performance and accurate environmental conditions, it is essential that all incubators are allowed to stabilise upon arrival and are professionally calibrated before first use. Calibration ensures the unit reaches and maintains the desired temperature, humidity, or CO₂ levels reliably.
Please ensure:
- The unit is left for a minimum of 12 hours upright and unpowered after delivery.
- Calibration is completed in accordance with manufacturer specifications before any use.
Failure to calibrate prior to use may result in inaccurate conditions and affect experimental integrity or compliance with laboratory standards.