BioQuell Microflow Class 2 Safety Cabinet Lab - Spares/Repair Lab

Item Code: ITEM-48802

Condition: Spares or Repairs

Location: Wales

NOTE - THIS IS A SPARES/REPAIRS ITEM
£ 200.00 exc. VAT (£ 240.00 inc. VAT)
1 in stock
If you are interested in this item or would like to place an order, please get in touch with us at , or by phone on +44 29 2086 4661.
Make us an offer / ask us a question
Description

BioQuell Microflow Class 2 Safety Cabinet


This item was removed from a facility where it was surplus to requirements.


It is in good cosmetic condition and powers on, however, the display shows a “Battery fail” error, preventing further functional testing.


The internal lights are not working, and an alarm sounds upon startup, which may be related to the battery fault.


Due to these errors, it is listed as Spares/Repair.

Item Details

Property Value Tested value
Electrical Supply N/A 1-Phase
Plug Included? (1-Phase) N/A Yes - Standard UK 3-Pin Mains Plug (BS 1363)
PAT Test Result N/A Pass
PAT Test Date N/A 01-10-2025
Voltage N/A 240V
Year of Manufacture N/A 2001
Last Service Date N/A 22/01/2015
General Quality Checks
UV Lamp Fitted N/A No
Internal Lights Fitted (Not UV) N/A Yes
Internal Lights Tested (Not UV) N/A Yes
Internal Lights Test Result N/A Fail
Fan Switch Working N/A Yes
Fan Powers On N/A Yes
Airflow Observed N/A Yes
Airflow Measured N/A Yes
Measured Average Airflow Rate (M/S) N/A 0.39
Airflow Rate Result (Above 0.25M/S is a Pass) N/A Pass
Internal Power Sockets Fitted N/A Yes
Tested Using Socket Tester N/A Yes
Internal Sockets Test Result N/A Pass
Floor Stand Included N/A No
Display Screen Fitted N/A No
Electronic Window Fitted N/A No
Delivery

Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.

Additional terms and conditions

These items are for spares/repair only, may be defective/unsafe, and are not intended for use.

We retain records of original source decontamination for all Safety Cabinets and Fume Hoods. Decontamination can consist of either Fumigation or Filter Removal. We are unable to provide condition report on any filters still present, we are unable to confirm their type or use suitability and we are unable to accept responsibility for overall function of the filters.
Filters should be assessed on receipt of goods and replaced by the user prior to use

Return policy

Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.

NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.