Scanlaf Mars 900 Biohazard Safety Cabinet Class 2 Lab Spares/Repairs

Item Code: ITEM-31155

Condition: Spares or Repairs

Brand: Scanlaf

NOTE - THIS IS A SPARES/REPAIRS ITEM
£ POA
Not in stock
SOLD - Unfortunately this item is no longer available. However, we may have similar items available for sale. Please get in touch with us at , or by phone on +44 29 2086 4661.
Description

This Scanlaf Mars 900 Biohazard Safety Cabinet Class 2 was removed from a university lab where it was surplus to requirements.


The item powers on successfully, with both the white and UV lights functioning. However, it does not achieve the necessary airflow range during testing, so it is listed for Spares/Repair.


The up button on the window is also intermittent.


It is in reasonably good cosmetic condition with some marks from use.


Working area - 51 x 90cm

Item Details

Property Value Tested value
Electrical Supply N/A 1-Phase
Plug Included? (1-Phase) N/A Yes - Standard UK 3-Pin Mains Plug (BS 1363)
PAT Test Result N/A Pass
PAT Test Date N/A 28-06-2023
Voltage N/A 240V
Year of Manufacture N/A 2008
Last Service Date N/A 2022-12-14
General Quality Checks
UV Lamp Fitted N/A Yes
UV Lamp Switch Working N/A Yes
UV Powers On N/A Yes
UV Measured N/A No
Internal Lights Fitted (Not UV) N/A Yes
Internal Lights Tested (Not UV) N/A Yes
Internal Lights Test Result N/A Pass
Fan Switch Working N/A Yes
Fan Powers On N/A Yes
Airflow Observed N/A Yes
Airflow Measured N/A Yes
Measured Average Airflow Rate (M/S) N/A 0.21
Airflow Rate Result (Above 0.30M/S is a Pass) N/A Fail
Internal Power Sockets Fitted N/A Yes
Tested Using Socket Tester N/A Yes
Internal Sockets Test Result N/A Pass
Floor Stand Included N/A No
Display Screen Fitted N/A Yes
Display Screen Working Correctly N/A Yes
Electronic Window Fitted N/A Yes
Electronic Window Operating Correctly N/A No
Electronic Window Fault Details N/A The up button is intermittent.
Delivery

Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.

Additional terms and conditions

These items are for spares/repair only, may be defective/unsafe, and are not intended for use.

We retain records of original source decontamination for all Safety Cabinets and Fume Hoods. Decontamination can consist of either Fumigation or Filter Removal. We are unable to provide condition report on any filters still present, we are unable to confirm their type or use suitability and we are unable to accept responsibility for overall function of the filters.
Filters should be assessed on receipt of goods and replaced by the user prior to use

Return policy

Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.

NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.