Description
Thermo Fisher HeraSafe HS9 Class 2 Biological Safety Cabinet
This item was removed from a facility where it was surplus to requirements.
The control panel is currently detached, leaving cables exposed. As a result, the unit failed the visual PAT inspection. It appears that the necessary parts required to reattach the control panel are present with the item.
Additionally, the glass window has a significant crack in the top right-hand corner.
Due to the above issues, the unit is being listed as Spares/Repair.
Item Details
| Property | Value | Tested value |
|---|---|---|
| PAT Test | ||
| Electrical Supply | N/A | Not applicable |
| PAT Test Result | N/A | Fail |
| PAT Test Fail Reason | N/A | Visual |
| Reason for Visual Fail | N/A | Enclosure |
| PAT Test Date | N/A | 25-08-2026 |
| Recertification Initial Checks | ||
| Overall Cosmetic Condition | N/A | Good |
| Stand Condition | N/A | N/A |
| Work Area Condition | N/A | Pass |
| Recertification System Checks | ||
| Display Screen | N/A | N/A |
| Service Details | ||
| Last Service Date | N/A | 10/8/21 |
| Other Details | ||
| Year of Manufacture | N/A | 2006 |
| Floor Stand Included | N/A | No |
Delivery
Free shipping is not available on this product. This item will be shipped by pallet delivery. Delivery will be made to kerbside only, unless otherwise confirmed in writing. Please contact us for a domestic or overseas shipping quote.
Additional terms and conditions
These items are for spares/repair only, may be defective/unsafe, and are not intended for use.
We retain records of original source decontamination for all Safety Cabinets and Fume Hoods. Decontamination can consist of either Fumigation or Filter Removal. We are unable to provide condition report on any filters still present, we are unable to confirm their type or use suitability and we are unable to accept responsibility for overall function of the filters.
Filters should be assessed on receipt of goods and replaced by the user prior to use
Return policy
Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.
NOTE: No cables, remotes, accessories, power supplies, external PCs, software media, hardkeys, consumables or any other items are included unless shown in the item photos or detailed above.