Kendro HERACell 240 Laboratory CO2 Incubator Lab

Item Code: ITEM-24694

Condition: Used

Brand: Kendro

Location: Wales

£ POA
Not in stock
SOLD - Unfortunately this item is no longer available. However, we may have similar items available for sale. Please get in touch with us at , or by phone on +44 29 2086 4661.
Description

This Kendro HERACell 240 Laboratory CO2 Incubator was removed from a university lab where it was surplus to requirements.


The unit powers on and is in good cosmetic condition with some minor usage marks.


The unit powers up and has been tested with a digital thermometer (not included) while charged with 5% CO2. The readout appeared to display 5% CO2 as would be expected.



When set to 25˚C the display showed 25.1˚C and the thermometer measured 26.4˚C.

When set to 37˚C the display showed 37.1˚C and the thermometer measured 40˚C.

When set to 55˚C the display showed 55.1˚C and the thermometer measured 62.3˚C.



Shelving:

2x shelves, they are both in good cosmetic condition.


Interior dimensions:

W62cm x H67cm x D53cm


Item Details

Property Value Tested value
Type Floorstanding
Heating Method Air Jacket
Refrigeration No
Minimum Temperature (ºC) 3 Above Ambient
Maximum Temperature 55
CO2 Sensor Type TC Sensor
With Hot Air Sterilisation Yes
Humidifier Automated RH Control
With UV Lamp No
Physical Specifications
Interior Volume (L) 238
Delivery

Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.

Additional terms and conditions

Incubators may be transported via a range of delivery methods, including palletised courier networks or direct delivery to the intended location of use. During transit, internal sensors, fans, or control systems may be affected by vibration or movement.

To ensure optimal performance and accurate environmental conditions, it is essential that all incubators are allowed to stabilise upon arrival and are professionally calibrated before first use. Calibration ensures the unit reaches and maintains the desired temperature, humidity, or CO₂ levels reliably.

Please ensure:
- The unit is left for a minimum of 12 hours upright and unpowered after delivery.
- Calibration is completed in accordance with manufacturer specifications before any use.

Failure to calibrate prior to use may result in inaccurate conditions and affect experimental integrity or compliance with laboratory standards.

Return policy

Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.

NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.