This VWR AP07R-20-V12E Refrigerated Circulator Chiller was removed from a university lab where it was surplus to requirements.
The unit is in good used cosmetic condition but does have various marks and signs of general use.
It powers up and has been tested against a digital thermometer (not included). it was tested at 37C and 70C which it reached and held. It was then tested to -20C and reached -19.
Also tested down to the minimum of -20degC and it reached -19degC.
The unit service warning does show it requires a filter and fluid change but other than that it seems to be fully operational.