Mettler Toledo ReactiR iC10 w FTIR Spectrometer System & Software Lab Fulty

Item Code: ITEM-25007

Condition: Spares or Repairs

NOTE - THIS IS A SPARES/REPAIRS ITEM
£ POA
Not in stock
SOLD - Unfortunately this item is no longer available. However, we may have similar items available for sale. Please get in touch with us at , or by phone on +44 29 2086 4661.
Description

This Mettler Toledo ReactiR iC10 w FTIR Spectrometer System was removed from a university lab where it was surplus to requirements.


We had the system inspected by a qualified engineer who reported the following;


They tested the FTIR instrument using the IC IR Software loaded on the Laptop which came with it.


The system powered on in a normal way but they couldn't get any signal (the signal level is zero). A warning notification appeared on the screen to say the system is misaligned. Scanning the background came up with too many velocity scan errors/velocity warnings which led to the failure of the scan and data was not saved accordingly.


They took off the fibre optic assembly and The MCT detector with the pre-amplifier to access the interferometer and check the components. The laser and laser detector is working fine and it guides the Michaelson interferometer at the right rate. The ZnSe beam splitter looks fine.


The IR Source (SiC source) hot spot is not centric and tends to be on one side. This may be due to the fact that the emitted light is taken from the side of the so housing. No indication of the flip of the polarity of the source each 15 min. A silicon oxide layer has developed on the source and they have tried to remove it as much as they can.


A MCT detector required liquid Nitrogen for cooling, they have switched to the DTGS detector to run the scan but even with that, it hasn't got any signal.


They expect one of the two reasons for the no signal issue is that either the system misaligned or the maiiboard is not detecting any signal from these two detectors. To exclude the misalignment reason they have moved the source in various directions but there is no indication of the change in the signal level. So most likely the main board has an issue with the signal processing.


It is sold for spare/repairs.

Delivery

The price includes free economy delivery (3-5 working days) to Mainland UK Postcodes EXCLUDING the following postcodes:

AB31-38, AB40-52, AB54-56, AB63 IV1-28, IV30-32, IV36-56, IV63 KA27-28 KW1-17 PA20-38, PA41-49, PA60-78 PH4-44, PH49-50 PO30-41 TR21-25 BT, GS, GY, HY, IM, JE, TDCU, ZE

If you live in one of the above postcodes please enquire for a shipping quote. Your item will be carefully packed and dispatched within 24 hours (Mon–Fri) of receiving payment.

International buyers please contact us for an overseas shipping quote.

Additional terms and conditions

These items are for spares/repair only, may be defective/unsafe, and are not intended for use.

Return policy

Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.

NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.