Bruker Ultraflex III LRF200 - CID Mass Spectrometer Lab Spares/Repairs

Item Code: ITEM-30225

Condition: Spares or Repairs

NOTE - THIS IS A SPARES/REPAIRS ITEM
£ 1,625.00 exc. VAT (£ 1,950.00 inc. VAT)
1 in stock
If you are interested in this item or would like to place an order, please get in touch with us at , or by phone on 029 2086 4661.
Make us an offer / ask us a question
Description

This Bruker Ultraflex III LRF200 - CID Mass Spectrometer was removed from a university lab where it was surplus to requirements.


The unit powers up but has a fault either with the MALDI source applied voltage, or the power supplies. The voltage on in the ion source plate is not reaching the set point of 25kV and is giving an error.

 

This could be

1. Source contamination. The source has been cleaned and will be cleaned again, but it only takes a trace of deposited carbon to cause a discharge. A thorough clean may fix it or a replacement source block.

2. High voltage power supply issue


The item is in reasonable, used cosmetic condition but the left-hand side panel/door is out of line so doesn't close properly. Both front and back bottom panels were removed to adjust the feet to move the unit around on its wheels.


It is sold for spares/repairs due to the fault.


Delivery

Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.

Additional terms and conditions

These items are for spares/repair only, may be defective/unsafe, and are not intended for use.

Return policy

Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.

NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.