Bruker Ultraflex TOF/TOF Mass Spectrometer Lab Spares/Repairs
Item Code: ITEM-38975
Condition: Spares or Repairs
Brand: Bruker
NOTE - THIS IS A SPARES/REPAIRS ITEM
£ 3,925.00 exc. VAT (£ 4,710.00 inc. VAT)
If you are interested in this item or would like to place an order, please get in touch with us at , or by phone on +44 29 2086 4661.
Description
Bruker Ultraflex TOF/TOF Mass Spectrometer
This was removed from a university lab where it was surplus to requirements.
The unit powers up but has not been tested further at our facility.
The unit has some faults as confirmed by the lab
On arrival system was in fault mode with 'laser interlock' error. Restarted the laser through marathon control and tested. The laser started to fire but then the high voltage switched off. Tested HV supplies and found that the ISS could only reach 19kv instead of the required 25kv. Suddenly the system started working again and peaks were seen on the sample BUT the system has 2 issues that are intermittent.
1. 30kv supply
2. Laser
Item Details
Property | Value | Tested value |
---|---|---|
Electrical Supply | N/A | 1-Phase |
Last Service Date | N/A | 2023-02-08 |
Delivery
Free shipping is not available on this product. This item will be shipped by pallet delivery. Please contact us for a domestic or overseas shipping quote.
Additional terms and conditions
These items are for spares/repair only, may be defective/unsafe, and are not intended for use.
Return policy
Buy with confidence. If you wish to return the item, for whatever reason, you may do so within 45 days subject to our terms and conditions. International buyers will have to pay return shipping.
NOTE: No cables, remotes, accessories, power supplies, consumables or any other items are included unless shown in the item photos or detailed above.