Items 21 to 37 of 37

Image of item

£ 1,150.00 (ex. VAT) 1 in stock

Barnstead Lab-Line MaxQ 4000 Incubator Shaker Model - SHKE4000-1CE

ITEM-28143

This Barnstead Lab-Line MaxQ 4000 Incubator Shaker  Model - SHKE4000-1CE was removed from a university lab where it was surplus to requirements.


The unit powers up and is working. It has been assessed by an engineer who replaced the lid struts and fitted a new drive belt.


The only issue is the main power button on the front of the unit doesn't appear to work. This does not affect the operation of the unit at all. It just needs to be switched on and off at the main supply.


The incubator function was tested with a digital thermometer (not included).


The temperature was first set to 37°c. when reached temp the thermometer read 37.6°c.

The temperature was then set to its max temp of 60°c, when it reached temp, the thermometer read 58.8˚c.



Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 1,100.00 (ex. VAT) 1 in stock

Medline Scientific Incubated Shaker Model - IS-971RF Spares/Repairs

ITEM-20411

This Medline Scientific Incubated Shaker Model - IS-971RF was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included). It was set to 37˚C but only reached 31˚C. The display would keep reverting to 22.2˚C as well. There was a note from the lab stating that it had some heating issues.


Inside dimensions - L89.5 x W63 x D48cm

Platform size - 46 x 76cm


The shaker speed ran up to 160rpm. We tried setting it to the max of 300 but it wouldn't run at that.


Cosmetically it is in a reasonable used condition but all the light tubes work bar one.


It is sold for spares/repairs due to the fault.

Image of item

£ 1,070.00 (ex. VAT) 1 in stock

New Brunswick Scientific G-25 Foor Standing Incubator Shaker Lab

ITEM-35604

New Brunswick Scientific G-25 Foor Standing Incubator Shaker


This was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included). It is in full working order.


It is in reasonable cosmetic condition but has seen some heavy use and has lots of marks and scuffs.


Shaker platform - 76 x 46cm

Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 1,000.00 (ex. VAT) 1 in stock

Lab Companion IS-971R Shaking Incubator Lab Cooling Fault

ITEM-39716

Lab Companion IS-971R Shaking Incubator


This was removed from a university lab where it was surplus to requirements.


The unit powers up but has a cooling fault. It is seemingly in otherwise working condition.


It is in good cosmetic condition with minor marks


It is sold for spares/repairs.

Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 985.00 (ex. VAT) 1 in stock

Thermo Scientific MAXQ 5000 SHKE5000-1CE (4351) Shaking Incubator Lab Faulty

ITEM-39911

Thermo Scientific MAXQ 5000 SHKE5000-1CE (4351) Shaking Incubator


This was removed from a university lab where it was surplus to requirements.


The system is faulty and requires a new controller.


It is in otherwise good cosmetic condition with minor marks.


It is sold for spares/repairs.

Image of item

£ 925.00 (ex. VAT) 1 in stock

Gallenkamp IOI400.XX2.C Floor Standing Shaking Incubator Lab

ITEM-12238

This Gallenkamp IOI400.XX2.C Refrigerated Shaking Incubator was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested.


It was tested using a Thermo Smart-Vue gauge (not included).

The gauge showed 37.4ºC when the unit was set to 37ºC.

The gauge showed 59.8ºC when the unit was set to 60ºC (Max).


The max rpm is 400, which the unit runs fine at. The first digit of the rpm screen does have issues, but the rpm can still be set ok.


It is not a refrigerated unit.


Inner dimensions: 85.5cm L x 35cm W.

Shaker plate dimensions: 69cm L x 49cm W.



Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 795.00 (ex. VAT) 1 in stock

Thermo Scientific Max Q 4000 Shaking Incubator Lab Spares/Repairs

ITEM-30287

This Thermo Scientific Max Q 4000 Shaking Incubator was removed from a university lab where it was surplus to requirements.


The unit would not power up in testing. We swapped the boards out but still could get it to power. There were loose parts inside, so we assume it has been stripped. It has a very worn drive section.


It's in reasonable cosmetic condition with some marks from use.


It is sold for spares/repairs.

Image of item

£ 730.00 (ex. VAT) 1 in stock

New Brunswick Scientific Innova 4000 Benchtop Incubator Shaker Lab

ITEM-30218

This New Brunswick Scientific Innova 4000 Benchtop Incubator Shaker was removed from a university lab where it was surplus to requirements.


The unit powers up and runs up to 500rpm. It was tested with a digital thermometer (not included).


The temperature was first set to 37°c, when reached, the thermometer read 42.8˚c

The temperature was then set to 60˚c (max) when reached, the thermometer read 64.4˚c


It is in reasonable cosmetic condition with some marks from use.


Internal Dimensions:

D48.3 W50.2 H33.6cm



Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 645.00 (ex. VAT) 1 in stock

Thermo Scientific MaxQ 4000 Analog Incubator Shaker SHKA4000-1CE Faulty

ITEM-37029

Thermo Scientific MaxQ 4000 Analog Incubator Shaker


This was removed from a university lab where it was surplus to requirements.


The unit would not power up and tripped the electrics during testing. We believe the motor controller is faulty. The gas struts for the lid have also failed.


It is in otherwise good cosmetic condition with marks from use.

Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 600.00 (ex. VAT) 1 in stock

New Brunswick Scientific Innova 4230 Refrigerated Incubator Shaker Spares/Repair

ITEM-36992

New Brunswick Scientific Innova 4230 Refrigerated Incubator Shaker Spares/Repair


This was removed from a university lab where it was surplus to requirements.


The unit powers but requires a replacement compressor and recharge as well as a new drive belt.


It is sold for spares/repairs.



Image of item

£ 550.00 (ex. VAT) 1 in stock

Gallenkamp Cooled Orbital Shaking Incubator INR-250 Lab

ITEM-36449

Gallenkamp Cooled Orbital Shaking Incubator INR-250


This was removed from a university lab where it was surplus to requirements.


The unit powers up and is in good working order (digital thermometer not included).


It is in reasonable cosmetic condition but has marks from use.



Image of item

SPARES/REPAIR ITEM

SPARES/REPAIR ITEM

£ 500.00 (ex. VAT) 1 in stock

New Brunswick Scientific Innova 4230 Refrigerated Incubator Shaker Spares/Repair

ITEM-39813

New Brunswick Scientific Innova 4230 Refrigerated Incubator Shaker Spares/Repair


This was removed from a university lab where it was surplus to requirements.


The unit powers up and the incubation function works, but the unit does not shake (the thermometer used for testing is not included).


It is in reasonably good cosmetic condition with minor marks.


It is sold for spares/repairs.

Image of item

£ 315.00 (ex. VAT) 1 in stock

Gallenkamp Floor Standing Orbital Shaker Incubator - Cat No IOX400.XX1.C Lab

ITEM-23979

This Gallenkamp Floor Standing Orbital Shaker Incubator was removed from a university lab where it was surplus to requirements.


The unit powers up and the shaker is fully working. We tested the incubation with a digital thermometer (not included). It was tested at both 37 and 70˚C. Both temperatures were quickly achieved.


It is in reasonably cosmetic condition but does have marks from use.


Internal dimensions are: Length 75cm x Width 54cm x Height 34cm


Image of item

£ 225.00 (ex. VAT) 1 in stock

Gallenkamp Plus II Laboratory Incubator Cat No. IPL.075.FD1.C Lab

ITEM-24814

This Gallenkamp Plus II Laboratory Incubator Cat No. IPL.075.FD1.C was removed from a university lab where it was surplus to requirements.


The item is in reasonable cosmetic condition but has marks and stains from use.


The unit powers on fine and the interior shaker works as intended however the higher it is set on the dial the slower it goes and vice versa. The shaker cannot be operated at temperatures higher than 60˚C.


The unit was tested with a digital thermometer (not included)


When set to 37˚C the display showed 35.6˚C and the thermometer 37.3˚C.

When set to 60˚C the display showed 56.8˚C and the thermometer 58.7˚C.



We then removed the shaker and the oven was set to 108˚C, which is the highest we could get it to go on the dial.

It reached 98.6 according to the display, while the thermometer measured 104.0˚C.



The shaker is a Gallenkamp Shaker Platform Cat no. IPL000.0S1.9.

Image of item

£ POA 1 in stock

Sanyo Orbital Shaking Incubator IOX402.XX1.C

ITEM-40948

This item is new in stock. Please enquire for further details.

Image of item

£ POA 1 in stock

Thermo Scientific MaxQ 6000 Incubator Shaker Model SHKE6000

ITEM-39063

This item is new in stock. Please enquire for further details.

Image of item

£ POA 1 in stock

New Brunswick Scientific Innova 4330 Refrigerated Incubator Shaker

ITEM-39182

This item is new in stock. Please enquire for further details.

Items 21 to 37 of 37

Filters

Price (£)

to

Condition

Brand

Category filters

Type

CO2 Sensor Type

Heating Method

With Hot Air Sterilisation

Humidifier

Refrigeration

Minimum Temperature (°C)

to

Maximum Temperature (°C)

to

With UV Lamp

Physical Specifications

Configuration

Interior Volume (l)

to