All Products
Items 2421 to 2440 of 3034
£ 120.00 (ex. VAT) 1 in stock
Signal Model 821 Gas Divider Lab
ITEM-25270
This Signal Model 821 Gas Divider was removed from a university lab where it was surplus to requirements.
The unit powers up but the remote/local switch on the back is broken.
It is in reasonably good cosmetic condition but has some marks from use.
£ 120.00 (ex. VAT) 1 in stock
Mettler Toledo LabMax Automatic Lab Reactor + ATB VAFU 71/4B-7 Pump Lab
ITEM-07207
This Mettler Toledo LabMax Automatic Lab Reactor with ATB VAFU 71/4B-7 Pump Lab was removed from a university laboratory where it was surplus to requirements.
We were unable to power the unit on but it is in good cosmetic condition with light marks and scratches.
Included with it are a pump/stirrer, an emergency stop button, and some cables.
£ 120.00 (ex. VAT) 1 in stock
Oxford Cryosystems AD41 Dry Air Unit Lab
ITEM-31352
The unit powers on and is in good cosmetic condition
The fans are working, the dial is responsive to the unit and the compressor kicks in a few seconds after turning the dial from ( 0 ) to ( I )- RUN
The unit has 32176 hours.
We were unable to test it further.
£ 115.00 (ex. VAT) 1 in stock
The General Electric Company P.L.C. Meniscograph Solderability Tester 685
ITEM-21121
The General Electric Company P.L.C. Meniscograph Solderability Tester 685
This was removed from a university lab where it was surplus to requirements.
It is in good cosmetic condition but has some marks from use.
It powers up and the button inputs are responsive but it was not tested further.
£ 115.00 (ex. VAT) 1 in stock
Union Carbide Liquified Gas Container Liquid Nitrogen Dewar Lab
ITEM-29956
Union Carbide Liquified Gas Container Liquid Nitrogen Dewar
This was removed from a university lab where it was surplus to requirements.
It is in reasonable cosmetic condition but has lots of scratches. It has not been tested.
£ 115.00 (ex. VAT) 1 in stock
Promega ReliaPrep LV 32 HSM Heater / Shaker Magnet Lab A1711
ITEM-24573
This Premega Reliaprep A1711 LV 32 HSM Instrument Shaker was removed from a university lab where it was surplus to requirements.
The unit powers up and runs smoothly at all speeds up to 700rpm. The power supply works and powers the unit.
The heating feature was tested with a digital thermometer (not included).
Set temp at 37ºC our thermometer read 29.9ºC
Set temp at 80ºC our thermometer read 55.3ºC
The readings are not the most accurate as it was hard to get the probe in a place it could read accurately.
Included are;
2x Tube Racks
One of the racks has a slight dent in the centre of the unit from general use. Both racks have 32 x holes 29.24mm in diameter
£ 110.00 (ex. VAT) 2 in stock
Meiji Techno Fiber Optic Lite Source Model - FL150 Model no. MI-1504FMJ Lab
ITEM-24544
This Meiji Techno Fiber Optic Lite Source Model - FL150 was removed from a university lab where it was surplus to requirements.
The unit powers up but has no bulb so we were unable to test it further.
It is in reasonably good cosmetic condition with marks from use.
Please note that we have multiples of this unit in stock. You may not receive the one in the photos but will receive one in near-identical working and cosmetic condition. If you would like to see photos of the exact unit you are receiving before it is shipped, we would be happy to provide this.
£ 110.00 (ex. VAT) 1 in stock
Heraeus Laboratory Incubator Typ B 5060 EK-C02 Lab
ITEM-19378
This Heraeus Laboratory Incubator Typ B 5060 EK-C02 was removed from a university lab where it was surplus to requirements.
The unit powers up and was tested with a digital thermometer (not included). We set it to 37C which it reached after about an hour. We then set it to 50C (mac) which it hit but hovered around 49 and 50. It took about 2 hours to get to the max temperature after hitting 37.
It is in reasonable cosmetic condition but has some heavy marks from use. It is not damaged.
Unit dimensions - L76 x W90 x H79cm
Internal dimensions - D57 x W60 x H60cm
6 Shelves 51 x 27.5cm
NEW ITEM
NEW ITEM
£ 110.00 (ex. VAT) 1 in stock
80 x Portwest PS57 Blue Safety PPE Hard Hats - Brand New
ITEM-23957
80 x Portwest PS57 Blue Safety PPE Hard Hats
These were removed from a university where they were surplus to requirements.
These hats are all brand new and sealed in bags.
They fit head sizes from 55-62cm.
The date stamp of manufacture is August 2020.
£ 110.00 (ex. VAT) 1 in stock
Beko Laboratory Freezer Lab
ITEM-31046
This was removed from a university lab where it was surplus to requirements.
It is in full working order and has been tested with a digital thermometer down to -20˚C
It is in good cosmetic condition with minor marks from use.
Internal Dimensions - D40 W43 H66cm
2 x Shelves 38 x 42cm
Bottom Drawer - 41 x 23 x 16cm
Top 'Fast Freeze' section - 37 x 42 x 14.5cm
Ice Bank Tray - 40.5 x 25cm
1 x Ice Cube tray holds 12 cubes 19.5 x 7.5cm
£ 105.00 (ex. VAT) 1 in stock
Genomic Solutions Vacuum Pump 6035A080-02 Lab
ITEM-21143
This Genomic Solutions Vacuum Pump 6035A080-02 was removed from a university lab where it was surplus to requirements.
The unit powers up and the gauge moves and appears to work as far as we are able to tell.
it is in good cosmetic condition with some marks from use.
£ 105.00 (ex. VAT) 1 in stock
Perkin-Elmer Lambda 5 UV/VIS Spectrophotometer Lab
ITEM-20302
This Perkin-Elmer Lambda 5 UV/VIS Spectrophotometer was removed from a university lab where it was surplus to requirements.
The item is in reasonable cosmetic condition but it is very stained and has other signs of prior use.
The item audibly powered on but there was no other indication of it working bar the sound. It was not tested further.
Please note that a decontamination has occurred but not all chemical residue can be removed, this can be seen in the images.
As such we advised hand hygiene and personal protective equipment must be used as a precaution.
£ 105.00 (ex. VAT) 1 in stock
8 x Hand Crimping Tools & 1 x Un-Crimping Tool 20mm, 11mm, Agilent Lab
ITEM-24628
8 x Hand Crimping Tools & 1 x Un-Crimping Tool
These were removed from a lab where they were surplus to requirements.
Included are;
3x Agilent 20mm Crimpers
2x 11mm Crimpers
3 x Unknown size Crimpers
1x Un-Crimper
Accessories:
4x sealed packs of Aluminium crimping caps (quantity unknown) (20.8mm Dia)
8x sealed packs of crimping 6 prong discs
2x AL/Silicone (Batch 189738) 5 prong silicone crimping discs
1x bag sealed 1 pack of 100
1x bag opened 1 pack of 100 (not confirmed)
They were not tested.
£ 100.00 (ex. VAT) 1 in stock
Genomic Solutions Jun-Air Compressor DK-9400 Model 6 Lab
ITEM-21141
This Genomic Solutions Jun-Air Compressor DK-9400 Model 6 was removed from a university lab where it was surplus to requirements.
The unit powers up but has a damaged current switch detector on it. It is in otherwise good cosmetic condition.
We ran it up to 100psi and seems to work well.
Please note the unit has not been pressure or safety inspected so it must be inspected by a competent engineer prior to use.
£ 95.00 (ex. VAT) 1 in stock
Syngene GBOX CHEMI 16 Gel Documentation Imaging System Lab + Transilluminator
ITEM-22043
This Syngene GBOX CHEMI 16 Gel Documentation Imaging System was removed from a university lab where it was surplus to requirements.
The unit is in reasonable cosmetic condition but the door has scratches and marks as well as staining and tape residue. The interior has staining on the back panel.
The system does power on but some of the bulbs may need replacing.
£ 95.00 (ex. VAT) 1 in stock
DACS 8 Channel Volume Controller Eightch
ITEM-17909
This DACS 8 Channel Volume Controller Eightch was removed from a university AV department where it was surplus to requirements.
The unit powers on but has not been tested further. It is in good cosmetic condition but has marks from use. The volume control knob is also missing.
£ 95.00 (ex. VAT) 2 in stock
Purite Select BIO40/BP/IT Water Purification System Lab
ITEM-20125
This Purite Select BIO40/BP/IT Water Purification System was removed from a university lab where it was surplus to requirements.
The unit is in reasonable cosmetic condition but has some marks from use.
It powers up and the screen is clear. It will likely need servicing before it is used again.
The UV lamp hours are at 6000. The forecast change date is 14/12/23.
Filter replacement was due 22/01/21.
DI pack replacement was due 22/01/21.
It has not been tested further.
£ 95.00 (ex. VAT) 1 in stock
DACS 8 Channel (Eightch) Volume Controller
ITEM-17910
This DACS 8 Channel (Eightch) Volume Controller was removed from a university AV department where it was surplus to requirements.
The unit powers up but has not been tested further. It is in good cosmetic condition with marks from use.
£ 90.00 (ex. VAT) 1 in stock
Topcon Surveying Reflector Prism with Tripod
ITEM-25033
This Topcon Surveying Reflector Prism with Tripod was removed from a university engineering lab where it was surplus to requirements.
The unit is in reasonable cosmetic condition but the reflector has been dented/bent and has chips/scratches and the tripod has lots of scratches and its strap is broken.
The carry case has scuffs and stains from prior handling. The tripod's cap has pen marks and scuff marks.
The unit has not been tested.
£ 90.00 (ex. VAT) 1 in stock
Biotron Biometra TRIO Thermoblock Thermal Cycler PCR Lab
ITEM-22061
This Biotron Biometra TRIO Thermoblock Thermal Cycler was removed from a university lab where it was surplus to requirements.
The item is in reasonable cosmetic condition but there are stains/marks and scratches. The start/stop button sticks but does work.
The item powered up and was tested with a digital thermometer (not included). Blocks 1, 2 and 3 were tested individually. Please note there may be some variation in the positioning of our probe and also in our own thermometer as well.
Block 1:
When set to 37˚C, the display showed 37˚C and the thermometer measured 36˚C
When set to 99˚C, the display showed 99˚C and the thermometer measured 97.6˚C
Block 2:
When set to 37˚C, the display showed 37˚C and the thermometer measured 36.2˚C
When set to 99˚C, the display showed 99˚C and the thermometer measured 97.1˚C
Block 3:
When set to 37˚C, the display showed 37˚C and the thermometer measured 35˚C
When set to 99˚C, the display showed 99˚C and the thermometer measured 96.3˚C
Items 2421 to 2440 of 3034