Items 981 to 1000 of 2900

Image of item

£ 124.69 (ex. VAT) 1 in stock

Eppendorf Masterycler 5330 Thermal Cycler + 5530 Plus PCR DNA Lab Spares Repairs

ITEM-10344

This unit was removed from a lab during a clearout.


The top unit lid heater powers on, but the bottom unit does not.


Its difficult to get a temperature readout on the lid as it goes to timer mode before reaching temperature.


Unit sold as spares / repairs.



Image of item

£ 285.00 (ex. VAT) 1 in stock

Eppendorf Mini Spin Plus Benchtop Centrifuge 5453 14500rpm Lab

ITEM-07005

This unit has been tested up to 14,500 rpm and seems to run smoothly.


We completed the following tests as described below. There appears to be one or two potential issues but generally seems to work OK.


Tested at 5,000rpm for 60secs.

Tested at 7,500rpm for 60secs.

Tested at 10,000rpm for 60secs.

Tested at 12,000rpm for 60secs.

Tested at max 14,500rpm for 60secs.


Through testing there was a brief issue where it showed at error 68 when run at 14,500rpm.

Which is a drive error.

It was noisy at higher rpm and then showed the error.


After that we ran all the above rpm tests without issue.


Then there was a brief lid error, where it wouldn't open.

But it went away after a couple of minutes, and then it worked fine.


Image of item

£ 71.25 (ex. VAT) 1 in stock

Eppendorf MiniSpin Benchtop Lab Centrifuge Lab - Spares Repairs

ITEM-10889

This unit was removed from a lab during a clearout.


On first testing we were able to get the unit running.


But now when powered on the error 'LID' comes up - the lid is shut and won't open.


Unit sold as spares / repairs.

Image of item

£ 142.50 (ex. VAT) 1 in stock

Eppendorf Mixmate Microplate Shaker PCR Block System Lab - Spares Repairs

ITEM-08731

This unit was removed from a lab store room as surplus to requirements.


Unfortunalte the unit has error code 08. The manual says to turn on/off for 5 seconds and try again but this doesn't appear to work. The online manual suggests its an electronics error but does not specifically state what the error is.


Image of item

£ 1,012.50 (ex. VAT) 1 in stock

Eppendorf Model 5810R Refrigerated Centrifuge Lab Equipment

ITEM-12518

This centrifuge powers on but will not run. It shows 'Error 6' when trying to start.


'Error 6' shows in user manuals as mains power voltage being too low, frequency converter has been overloaded or that the Brake is faulty.

We have checked the power supply which seems okay so we assume the internal converter is faulty and/or the Brake is faulty.

It also will not hold cooled temperature. Max minus temp should be -9°C. This unit cooled down to -7°C but went straight back to +4°C and wouldn't cool back down again.


Due to the above this centrifuge is being sold as Spares or Repair only.

Image of item

£ 152.30 (ex. VAT) 1 in stock

EPSON EB-480 Bright Ultra Short Throw Projector Office

ITEM-09670

This unit was removed from an office as surplus to requirements. It is in full working order.

Image of item

£ 159.07 (ex. VAT) 1 in stock

Epson EB-485W Ultra Short Throw Projector Office

ITEM-09625

This unit was removed from an office as surplus to requirements. It is in full working order.

Image of item

£ 106.88 (ex. VAT) 2 in stock

Epson ELPLP76 Projector Spare Lamp L2705

ITEM-10630

This new / unused lamp was removed from a university store as surplus to requirements.

Image of item

£ 106.88 (ex. VAT) 1 in stock

Epson EMP-700 PowerLite Home Cinema Projector

ITEM-10203

This unit was removed from an office as surplus to requirements.


It appears to be in good working order.


Some cables are included but there does not appear to be a remote.

Image of item

£ 106.88 (ex. VAT) 1 in stock

Epson EMP-710 Super Ultra Portable (XGA) Projector With Remote & Cables

ITEM-10205

The unit was removed from a university department as surplus to requirements.


It appears to be in working order.

Image of item

£ 106.88 (ex. VAT) 1 in stock

Epson EMP-740 LCD Projector

ITEM-09957

This project was removed from an office as surplus to requirements.


It appears to be in working order with a clear image.


Lamp hours are 0 on high brightness and 398 on low brightness.



Image of item

£ 57.00 (ex. VAT) 1 in stock

Epson WorkForce GT-1500 Flat-Bed Scanner

ITEM-11843

This unit was removed from an office as surplus to requirements.


The unit was in full working order, but we have lost the power supply cable for it.

Image of item

£ 643.03 (ex. VAT) 1 in stock

EquiBio EASYJECT PLUS Electroporation System Single/Twin-Pulse with Accessories

ITEM-06376

EquiBio EASYJECT PLUS Electroporation System Single/Twin-Pulse with Accessories.  This system was pulled from a working Laboratory in a top UK University as part of a Lab Clearance.


The System consists of:


Main System on wheels plus the following accessories:


Remote Control Unit 

2 x Electroporation Chambers

Printer (the printer appears to have a plate missing from top of unit - please see photo's)

Cables


The Main Unit and Remote Control Unit both power on but we do not have the expertise to fully test the whole system.


 Please note:  We have not powered on or tested the accessories

Image of item

£ 180.00 (ex. VAT) 1 in stock

Erich Jaeger VD1000 Power Supply Type ML-Konstanter Lab

ITEM-12654

This unit was removed from a lab during a clearout.


We have powered the unit on but not tested it further.

Image of item

£ 64.30 (ex. VAT) 1 in stock

Ernst Leitz GMBH Wetzlar Transformer Power Supply

ITEM-09409

This unit was removed from a lab store-room as surplus to requirements.

We have powered it on but not tested it further.

Image of item

£ 67.69 (ex. VAT) 1 in stock

Ernst Leitz GMBH Wetzlar Transformer Power Supply

ITEM-09406

This unit was removed from a lab store-room as surplus to requirements.


We have powered it on but not tested it further.

Image of item

£ 135.38 (ex. VAT) 1 in stock

Erweka ZT3-1 Disintegration Tester Lab

ITEM-08702

This unit was removed from a lab store room as surplus to requirements. The unit powers on but has not been tested further.

Image of item

£ 285.00 (ex. VAT) 1 in stock

Eschmann TD411-RS Electrosurgical Unit Lab Medical

ITEM-09444

This unit was removed from a lab as surplus to requirements.


We have powered it on but not tested it further.

Image of item

£ 285.00 (ex. VAT) 1 in stock

ET3000 Electronic Tourniquet Scandmed Lab Medical

ITEM-11510

This unit was removed from a university life sciences lab as surplus to requirements.


We have powered on the unit but not tested it further.

Image of item

£ 128.25 (ex. VAT) 1 in stock

Eurisem Technics SE126 Digital Humidity / Temperature Meter

ITEM-10646

This unit was removed from a university lab as surplus to requirements.


The lab manager assured us it was working properly when last used.


We have only tested it to power-on ourselves.

Items 981 to 1000 of 2900