Items 661 to 680 of 3302

Image of item

£ 112.52 (ex. VAT) 1 in stock

Bio-Tek uQuant Microplate Spectrophotometer μQuant Lab Spares or Repair

ITEM-17848

This Bio-Tek uQuant Microplate Spectrophotometer μQuant was removed from a university lab where it was surplus to requirements.


The unit is in reasonably good, used cosmetic condition but it does have various marks and signs of use.


The unit powers up fine although it requires a 24volt PSU (not included).


It runs a self-test on startup which fails and shows '0201 error'. This is a Y-Axis error. We cannot get it to do anything else due to this error.


From the manual:


"Y-axis (carrier left/right) motor did not find the home sensor A motor was not able to move to its “home” position as registered by feedback from an optical sensor. Probable Causes: • Bearings on the Y-axis rails are dirty and worn and causing too much friction. • Defective or broken optical sensor. • Defective motor controller PCB or cable. • Carrier not able to move into read chamber. Check the carrier’s path for obstructions."


It is sold as Spares or Repairs


Image of item

£ 1,767.78 (ex. VAT) 1 in stock

BioCare intelliPATH FLX IPS000IINTL Automated Slide Stainer Pathology Lab

ITEM-13722

This system was removed from a hospital pathology lab as surplus to requirements.


When we arrived, the system was in situ on a benchtop. We were informed that it was working when last used, they simply do not use it anymore.


At our site, we have powered it on but not tested it further.


There is no PC, or software - it is just the unit. There are some waste bottles and other accessories as shown in the photos.

Image of item

£ POA 1 in stock

Bioer Life Touch Thermal Cycler Model - TC-96/G/H(b)B Lab

ITEM-19880

This Bioer Life Touch Thermal Cycler Model - TC-96/G/H(b)B was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included). We set a program to run at 37˚C, 70˚C, 99˚C and then back down to 70˚C, 37˚C and finally 10˚C. It heats and runs accurately.


It has a 96 Well block with 40mm wells. (0.2mL)


It is in good cosmetic condition with only light marks from use.


Image of item

£ POA 1 in stock

Bioer Technology ThermoE Cooling & Heating Block Model - HB-A100-9602 Lab

ITEM-20004

This Bioer Technology ThermoE Cooling & Heating Block Model - HB-A100-9602 was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included). Set temperature to 37˚c which it reached quickly (thermometer read 36˚c) Then set to max temp of 105˚c which it reached after about 5mins. We tried to set it to 15˚C but the cooling light didn't come on.


It is in good cosmetic condition but has some marks from use.


Block - 96 x 0.2ml



Image of item

£ 94.82 (ex. VAT) 2 in stock

Biofeedback Systems Conductivity Meter Model - SCL200 Lab

ITEM-17762

This Biofeedback Systems Conductivity Meter Model - SCL200 was removed from a university lab where it was surplus to requirements.


The unit powers on with 2 x 9V Batteries (not included). A sound was coming out of the speaker in the rear, but it was not tested further.


It is in reasonably good cosmetic condition but has some marks and discolouration from use.


Please note that we have multiples of this unit in stock. You may not receive the one in the photos but will receive one in near-identical working and cosmetic condition. If you would like to see photos of the exact unit you are receiving before it is shipped, we would be happy to provide this.

Image of item

£ 67.47 (ex. VAT) 2 in stock

Biofeedback Systems Model - SCL 100 Skin Conductivity Machine Lab

ITEM-17769

This Biofeedback Systems Model - SCL 100 Skin Conductivity Machine was removed from a university lab where it was surplus to requirements.


It powered on with a 9V PP9 battery (not included), but was not tested further. It is in reasonably good cosmetic condition with scuffs and marks from use.



Image of item

£ 65.62 (ex. VAT) 1 in stock

Biofeedback Systems Model EMG 100 Skin Conductivity Machine Lab

ITEM-17766

This Biofeedback Systems Model EMG 100 Skin Conductivity Machine was removed from a university lab where it was surplus to requirements.


It powers on with 9V PP9 Battery (not included) but was not tested further.


It is in reasonable cosmetic condition but is missing the back of the unit as well as one of the feet.


Image of item

£ 73.15 (ex. VAT) 2 in stock

Biofeedback Systems Model SCL200 Conductivity Meter Lab

ITEM-17763

This Biofeedback Systems Model SCL200 Conductivity Meter was removed from a university lab where it was surplus to requirements.


The unit powers on with a 9V Battery (not included). A sound was coming out of the speaker in the rear, but it was not tested further.


It is in good cosmetic condition but has some marks and discolouration from use.


Please note that we have multiples of this unit in stock. You may not receive the one in the photos but will receive one in near-identical working and cosmetic condition. If you would like to see photos of the exact unit you are receiving before it is shipped, we would be happy to provide this.

Image of item

£ 67.47 (ex. VAT) 1 in stock

Biofeedback Systems Skin Conductivity Machine Model - Biodata T100 Lab

ITEM-17764

This Biofeedback Systems Skin Conductivity Machine Model - Biodata T100 was removed from a university lab where it was surplus to requirements.


It powers on with a 9V PP9 Battery (not included) but was not tested further.


Cosmetically it is in reasonably good condition but has marks and scuffs from use.


Image of item

£ 75.87 (ex. VAT) 1 in stock

Biofeedback Systems Skin Conductivity Machine Model - BSL 758 SCL 100 Lab

ITEM-17767

This Biofeedback Systems Skin Conductivity Machine Model - BSL 758 SCL 100 was removed from a university lab where it was surplus to requirements.


It powers on but requires a 9V PP9 battery (not included)


Cosmetically it is in reasonable used condition with a crack on the front. It doesn't seem to affect the usage of the unit at all though.


Image of item

£ 69.45 (ex. VAT) 1 in stock

Biofeedback Systems Skin Conductivity Machine Model - EEG 100 Lab

ITEM-17765

This Biofeedback Systems Skin Conductivity Machine Model - EEG 100 was removed from a university lab where it was surplus to requirements.


It was tested with our new 9V battery (not included) and powers on. it was not tested further.


Cosmetically it is in reasonably good condition but has marks and scuffs from use.


Image of item

£ 58.52 (ex. VAT) 2 in stock

Biofeedback Systems Skin Conductivity Machine Model - SCL 100 Lab

ITEM-17768

This Biofeedback Systems Skin Conductivity Machine Model - SCL 100 was removed from a university lab where it was surplus to requirements.


It powers on with 9V PP9 Battery (not included) but was not tested further. It is in reasonably good cosmetic condition but has some marks and scuffs from use.


Please note that we have multiples of this unit in stock. You may not receive the one in the photos but will receive one in near-identical working and cosmetic condition. If you would like to see photos of the exact unit you are receiving before it is shipped, we would be happy to provide this.



Image of item

£ 67.65 (ex. VAT) 1 in stock

Biofeedback Systems Skin Conductivity Machine/Meter Model SCL 100 Lab

ITEM-17770

This Biofeedback Systems Skin Conductivity Machine/Meter Model SCL 100 was removed from a university lab where it was surplus to requirements.


The unit powers on with a 9V PP9 battery (not included). It has not been tested further.


It is in reasonably good cosmetic condition but the battery low bulb is missing.


Image of item

£ 128.66 (ex. VAT) 1 in stock

Biometra Gel Electrophoresis Power Supply Lab

ITEM-13715

This Biometra Power Supply was removed from a university where it was surplus to requirements.


The unit powers on but has not been tested any further. It will need a safety inspection before use.


The unit is in reasonable cosmetic condition, but has signs of wear as it was well used. The plastic on one edge of the knob is broken, but this does not hinder the function.


The original feet had decayed and have been replaced.




Image of item

£ 100.70 (ex. VAT) 1 in stock

Biometra Maxidry D64 Electrophoresis Gel Dryer Lab

ITEM-09862

This unit was removed from a lab as surplus to requirements and we were assured the unit was in working order prior to removal from the lab


However, we have only tested it to power-on ourselves.


Therefore we are happy to sell the unit on the basis that it is in working order, and if you encounter any problems we can offer a 90-day return to base warranty.


Within the UK, we will also arrange the return postage so you are not out of pocket if a problem does arise with the unit.

Image of item

£ 187.50 (ex. VAT) 1 in stock

Biometra Mini Hybridization Oven OV3 Lab

ITEM-19582

This Biometra Mini Hybridization Oven OV3 was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included).

When set to 37˚C our thermometer read 37˚C.

When set to 100˚C our thermometer measured 73˚C. It would not get any higher than this. 73 seems to be the max temp.


The rotisserie shaker seems to be working and all the buttons are responsive.


It is in good cosmetic condition with some minor marks from use.

Image of item

£ POA 1 in stock

Biometra Standard Power Pack P25 & CompactM Gel Electrophoresis Units Lab

ITEM-19847

Biometra Standard Power Pack P25 & CompactM Gel Electrophoresis Units


These were removed from a university lab where they were surplus to requirements.


Included are;


1 x Biometra Standard Power Pack P25

The unit powers on but has been labelled as 'Out of Order' by the lab it was collected from. We are unable to confirm the fault. It is in reasonably good cosmetic condition with marks.


2 x Biometra CompactM Electrophoresis Units

Order numbers

  1. 025-200
  2. 025-200

Both units show signs of use.


2 x Biometra CompactM Casting Systems

Order numbers

  1. 025-210
  2. 025-210

Image of item

£ POA 1 in stock

Biometra T1 Thermocycler T1-Thermoblock Thermal Cycler PCR Lab

ITEM-19346

This Biometra T1 Thermocycler T1-Thermoblock Thermal Cycler was removed from a university lab where it was surplus to requirements.


The unit powers up and has been tested with a digital thermometer (not included).


It was set to 37˚C and our thermometer read 36˚C. It was then set to the minimum of -0.3˚C and reached 0. Finally, it was set to the max of 99.9˚C and reached 103˚. There was a label on it when it came to us saying it doesn't cool down, but we were unable to confirm any faults. It seems to heat and cool well but we have not run any cycles with it.


It is in good cosmetic condition with some marks from use.


Image of item

£ 642.83 (ex. VAT) 1 in stock

Biometra T3 Bench-Top Thermocycler 3x 48 Well Blocks Lab PCR

ITEM-15325

This Biometra T3 Bench-Top Thermocycler was removed from a laboratory where it was surplus to requirements.


The unit powers on and has been tested. All 3 blocks heat as they should. They have all been tested to 37 Celsius and the max of 99 Celsius.


There are some cosmetic faults however. 2 of the block lids are faulty. The centre and far right lid have broken top controls. The lid heater can be adjusted but the top knobs are snapped off and have just been placed bakc on.


However the unit does appear to work and all 3 blocks heat as they should.


The system has not been tested with samples.




Image of item

£ 546.41 (ex. VAT) 1 in stock

Biometra T3 Thermoblock Benchtop Thermal Cycler Lab PCR

ITEM-15324

This Biometra T3 Thermoblock Bench-Top Thermal Cycler was removed from a laboratory where it was surplus to requirements.


The unit powers on and in testing was found to be working, but there are a few issues with it. The screen is quite dim and difficult to read and the unit takes longer than it should to boot up. Some of the text at the bottom of the screen is also scrambled.


All 3 heating blocks appear to be in full working order and were tested at both 37 celsius and the max of 99 celsius.


Unit has not been tested with samples.